Content Status

Type

Content

There could be overall weak or no signals on the strip, including or except, the Conjugate Control (CC) zone.

Figure: LPA showing No Signal Except CC 

 

 

 

Reasons and Troubleshooting for No Signal Except CC are given in the table below:

 

Table: Reasons and Troubleshooting for No Signal Except CC

REASON FOR ERROR

SOLUTION

The quality of extracted DNA does not allow efficient amplification 

Repeat the extraction step​

Amplification Mixes (AM-A and AM-B) were either not mixed properly or were added in the wrong amounts​

Prepare a new master mix and repeat the test​ 

The incubation temperature was too high​

Repeat reverse hybridization​

The extracted bacterial species cannot be detected by the Universal Control

Use an alternative identification method

 

Resources 

 

 

Kindly provide your valuable feedback on the page to the link provided HERE
 

Content Creator

Reviewer